Smart LAC multi-spot measurement
Smart LAC: H410 multi-spot measurement
Notes for external light source incident measurement

What is external light source incident measurement…

Smart LAC H410 can measure the angle of the light source prepared by the user.
If you select "External" in the image processing section, the light source inside the sensor head is turned off, enabling measurement of an external light source.

Usage example
Notes

The H410 series can measure wavelengths from 450 to 940 nm.
When shipped from our company, it is calibrated according to the wavelength of the light source inside the sensor head.
Therefore, when measuring at a wavelength different from the light source wavelength, an error occurs in the angle of the measurement result due to the effect of chromatic aberration.
Regarding the angle error, we will prepare the data as a reference value.
Please contact us to request the wavelength and angle error data you are considering using.

Spec List
H410-175B1 1.75 1 450±10 Semiconductor laser 450±10nm φ1mm(Fixed) ±0.50deg : 0~300mm
±1.00deg : 0~200mm
±1.75deg : 0~120mm
(Varies depending on measurement range)
Less than 0.39mW (Class 1) 0.14kg(Excluding AC Adaptor) HB10 HPU-1000 and H410 are required for measurement. 2D 3D
H410-175G1 1.75 1 520±10 Semiconductor laser 520±10nm φ1mm(Fixed) ±0.50deg : 0~300mm
±1.00deg : 0~200mm
±1.75deg : 0~120mm
(Varies depending on measurement range)
Less than 0.39mW (Class 1) 0.14kg(Excluding AC Adaptor) HB10 HPU-1000 and H410 are required for measurement. 2D 3D
H410-175R 1.75 1 655±10 Semiconductor laser 655±10nm φ1mm(Fixed) ±0.50deg : 0~300mm
±1.00deg : 0~200mm
±1.75deg : 0~120mm
(Varies depending on measurement range)
Less than 0.39mW (Class 1) 0.14kg(Excluding AC Adaptor) HB10 HPU-1000 and H410 are required for measurement. 2D 3D
H410-175S1 1.75 1 852±10 Semiconductor laser 852±10nm φ1mm(Fixed) ±0.50deg : 0~300mm
±1.00deg : 0~200mm
±1.75deg : 0~120mm
(Varies depending on measurement range)
0.78mW or less (Class 1) 0.14kg(Excluding AC Adaptor) HB10 HPU-1000 and H410 are required for measurement. 2D 3D